Bug #24
Extra Data Packets
Start date:
01/15/2019
Due date:
% Done:
0%
Estimated time:
Description
While running a Comprehensive Performance Test (CPT) for the record, while installed on the Spacecraft (at LASP), the instrument exhibited an anomalous behavior. The instrument collects the incorrect number of data packages (extra science packets) relative to the issued command. This has been intermittent, sometimes acquiring 1 extra packet, sometimes 2 extra packets, and sometimes no extra packets. Anomaly was also noticed when using the EMIRS_ACQ command (in Section 4.3.1, step b). Anomaly was also noticed during Aliveness testing.
History
#1 Updated by Igor Lazbin almost 6 years ago
- Status changed from New to In Progress
#2 Updated by Igor Lazbin over 5 years ago
- Status changed from In Progress to Resolved