Description
While running a Comprehensive Performance Test (CPT) for the record, while installed on the Spacecraft (at LASP), the instrument exhibited an anomalous behavior. The pointing mirror assembly performed an unexpected rotation after the SC Status Watchdog was enabled (Section 4.3.1, Step aaa). Similar behavior was exhibited during aliveness testing, after SC Safe Bit was set (=1, safed), while the mirror was at home.
History
#1 Updated by Igor Lazbin almost 6 years ago
- Status changed from New to In Progress
#2 Updated by Igor Lazbin over 5 years ago
- Status changed from In Progress to Resolved